由于缺少路径、多个磁盘托架出现故障
适用场景
问题描述
- 多个磁盘在不同抽盒的磁盘架中显示出现故障
- 具有相同IOM的不同抽盒的路径在的输出中显示错误
::> system node run -node <node name> -command storage show fault -v
Enclosure:
Element Status Status Bytes Status Descriptions
1: OK 01,00,02,00 FAIL
Drawers Control Module:
Element Status Status Bytes Status Descriptions
1 [IOM12B A] : OK 01,00,01,01 REPORT, TYPE BIT0
2 [IOM12B A] : OK 01,00,01,01 REPORT, TYPE BIT0
3 [IOM12B A] : OK 01,00,01,01 REPORT, TYPE BIT0
4 [IOM12B A] : OK 01,00,01,01 REPORT, TYPE BIT0
5 [IOM12B A] : OK 01,00,01,01 REPORT, TYPE BIT0
6 [IOM12B B] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
7 [IOM12B B] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
8 [IOM12B B] : OK 01,00,01,01 REPORT, TYPE BIT0
9 [IOM12B B] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
10 [IOM12B B] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
- 重新拔插IOM后、磁盘似乎工作正常、但报告新的IOM错误:
Drawers Control Module:
Element Status Status Bytes Status Descriptions
1 [IOM12B A] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
2 [IOM12B A] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
3 [IOM12B A] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
4 [IOM12B A] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
5 [IOM12B A] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
6 [IOM12B B] : OK 01,00,01,01 REPORT, TYPE BIT0
7 [IOM12B B] : OK 01,00,01,01 REPORT, TYPE BIT0
8 [IOM12B B] : OK 01,00,01,01 REPORT, TYPE BIT0
9 [IOM12B B] : NONCRITICAL 03,00,11,01 REPORT, PATH FAIL, TYPE BIT0
10 [IOM12B B] : OK 01,00,01,01 REPORT, TYPE BIT0